pH Measurement FAQs
- If I order a pH meter, what accessories do I need to use with
it?
- My coworker is using an ORP electrode to measure the same
solution as I, but our readings are not even close. Could
there be something wrong with my electrode?
- Can I measure the pH of a gas?
- How should I store my electrode?
- What is the difference between a combination electrode and a
sensing electrode with a reference cell?
- How does one take soil pH measurements?
- Why is a double junction electrode better than a single junction
electrode?
- Is automatic temperature compensation (ATC) really necessary?
- When do you use a half cell?
- How often should I calibrate my pH meter?
- How can you unclog a pH electrode? How can you restore a
dray pH electrode?
- How fare can my pH electrode be from my meter? What if it
is too far?
- How do I extend the ATC wire on my industrial pH electrodes?
- What is dead band?
- If measureing samples across the entire pH range, what slope
should be used?
- What is the difference between blue glass and amber glass, and
what does that have to do with pH measurement?
- If I order a pH meter, what accessories do I need to use with
it?
You need a pH electrode and at least two pH buffers, on a pH 7 and the other at either pH 4 or 10.
Go to Top
- My coworker is using an ORP electrode to measure the same
solution as I, but our readings are not even close. Could
there be something wrong with my electrode?
No. Because ORP is a relative measurement, it is almost impossible to compare two ORP electrodes directly.
ORP electrodes come equipped with bands made up of platinum, gold, or hydrogen, fore example. Each band type
will give you a different reading in the same solution. Even if the electrodes are of the same band type, the
leak rate through the reference junction will affect your readings.
Instead, simply measure two solutions and note the difference between the two electrodes. Once again, the
difference between two solutions should compare. You should be looking for a change of state, rather than an
absolute measurement. You can check your electrode using pH buffer and quinhydrone.
Go to Top
- Can I measure the pH of a gas?
The only way to measure the pH of a gas is to dissolve it into distilled water and measure the mixture. Technically,
the pH of the distilled water/gas misture will be that of the gas.
Go to Top
- How should I store my electrode?
The best solution fo electrode storage is 4M KCL. pH 4 buffer, pH 7 buffer, or tap water are also acceptable. Never
store you electrode in distilled water.
Go to Top
- What is the difference between a combination electrode and a
sensing electrode with a reference cell?
A combination electrode is more convenient and requires a smaller sample container and
volume. The sensing electrode with reference cell combination allows you to select the
reference cell most compatible with your solution. You can select the double junction,
calomel, or ROSS SURE FLOW half cell. You will probably get better life from this
combination, and can replace each cell individually.
Go to Top
- How does one take soil pH measurements?
Use our soil electrodes EW-05992-62 or EW-05992-60. Prepare the sample by
combining a 5 g soil sample with 5 g of distilled water, mixing thoroughly, and allowing
the mixture to settle for 10 minutes. Carefully insert probe so bulb is in the soil part and
the junction is in the supernatant. Allow reading to stabilize.
Go to Top
- Why is a double junction electrode better than a single junction
electrode?
A double junction electrode is less likely to become clogged because the second junction
is located higher up in the probe out of contact with the fluid.
Go to Top
- Is automatic temperature compensation (ATC) really necessary?
The necessity of ATC depends on the required accuracy of a pH reading. pH readings
vary with temperature. For example, a sample with a pH of 7 at 25°C, may have a pH of
7.08 at 5°C and a pH of 6.98 at 60°C.
Go to Top
- When do you use a half cell?
When you are measureing a pressurized flow in a stream or pipe. The reference half cell
would be mounted upstream, the measureing electrode would be mounted downstream.
Go to Top
- How often should I calibrate my pH meter?
Before each use or set of uses.
Go to Top
- How can you unclog a pH electrode? How can you restore a
dray pH electrode?
First check the interior wire. If corrosion is evedent, replace the electrode. If not, then
soak the electrode in pH 4 buffer solution at 50°C for 2-4 hours. Restore a dry
electrode by soaking it in tap water after rinsing out the refill chamber with distilled
and refilling with the proper solution.
Go to Top
- How fare can my pH electrode be from my meter? What if it
is too far?
200 feet (61 meters) is the maximum distance an electrode can be from a pH meter. If the
distance is greater, you will need a transmitter. Use either a 4 to 20 mA transmitter (make
sure your pH meter accepts 4 to 20 mA signal) or purchase an industrial electrode with a
built-in transmitter. A transmitter will allow you to use your electrode up to 1000 feet
(305 meters) from you meter provided you are not in a noisy environment.
Go to Top
- How do I extend the ATC wire on my industrial pH electrodes?
You can extend the ATC wire with copper wire or a special extension cable.
Go to Top
- What is dead band?
When the input is varied to a device until a detectable output charge is observed then the
input is varied in the opposite direction until a detectable output charge is seen, the
increment through which the input signal is varied is the deadband.
Go to Top
- If measuring samples across the entire pH range, what slope
should be used?
The upper end or pH 10 buffer.
Go to Top
- What is the difference between blue glass and amber glass, and
what does that have to do with pH measurement?
The valance of Na+ is much large than H+. Amber glass has a smaller pore size thus
possibly discriminating between H+ and Na+ allowing only the smaller H+ to enter the
greatly eliminating Na+ interference.
Go to Top
BVA Scientific * 11914 Radium Dr * San Antonio, TX 78216 * 1-800-257-6474
Should you have any questions, please feel free to contact us and one of our representatives will get back to you shortly.
We accept Visa/MasterCard. For account set-up
information and our terms and conditions, click here.
Copyright© 2007 ALL RIGHTS RESERVED